The feasibility of stimulated emission depletion( STED) microscopy using a solid immersion lens was investigated. First, the theoretical feasibility of the considered system is discussed based on a vectorial field algorithm that uses a stratified medium composed of a SIL air-gap and test sample. Using the simulation, we verified that evanescent waves with much higher spatial frequencies corresponding to the high numerical aperture in the air-gap can be utilized to achieve a higher resolution than a confocal fluorescent image without a depletion beam. The simulated expectation was supported by actual imaging on two types of samples: fluorescent beads with a 20 nm diameter and an actin sample with a filamentous structure. The lateral resolution of the system was determined to be 34 nm via the imaging results on the nano-beads. The system was quite promising for achieving nano-scale surface imaging of biological samples; an even higher resolution was achieved by adjusting the wavelength and the intensity of the depletion beam. (C) 2017 Optical Society of America
키워드
- 제목
- Investigation on achieving super-resolution by solid immersion lens based STED microscopy
- 저자
- Kim, Wan-Chin; Moon, Hyungbae; Lee, Won-Sup; Lim, Geon; Choi, Guk-Jong; Kang, Donyoung; Lee, Hyungsuk; Park, No-Cheol
- 발행일
- 2017-07
- 유형
- Article
- 저널명
- Optics Express
- 권
- 25
- 호
- 14
- 페이지
- 16629 ~ 16642