Real-time in vivo monitoring of intraocular pressure distribution in the anterior chamber and vitreous chamber for diagnosis of glaucoma
  • 서호철
  • Hong Y.-M.
  • 정원기
  • Park W.
  • Lee J.
  • 외 4명
Citations

WEB OF SCIENCE

0
Citations

SCOPUS

24

초록

Glaucoma causes irreversible vision loss due to optic nerve damage and retinal cell degeneration. Since high intraocular pressure (IOP) is a major risk factor for glaucoma development, accurate IOP measurement is crucial, especially intravitreal IOP affecting the optical nerve and cells. However, conventional methods have limits in selectively and directly detecting local retina pressure. Here, we present continuous measurements of local IOP values in the anterior chamber and vitreous chamber of living animals using minimally invasive probes with pressure-sensitive transistors. After inducing glaucoma in animal models, we compared the local IOP distribution between normal and glaucomatous eyes. We also compared IOP values detected in the cornea using tonometry measurements. Our findings revealed that glaucoma induced higher IOP in the vitreous chamber than in the anterior chamber, indicating that measuring IOP in the vitreous chamber is key to the glaucoma model. This progress offers future directions for diagnosis and treatment of glaucoma.

제목
Real-time in vivo monitoring of intraocular pressure distribution in the anterior chamber and vitreous chamber for diagnosis of glaucoma
저자
서호철Hong Y.-M.정원기Park W.Lee J.Kim H.K.Byeon S.H.Kim D.W.Park J.-U.